High-Sensitivity Cardiac Troponin Rule-Out Pathways: Assay-Specific Thresholds and Where They Do Not Transfer
pdf

Keywords

High-sensitivity troponin
Myocardial infarction
Rule-out pathway
Assay verification
Change control

Abstract

Rapid rule-out pathways for myocardial infarction using high-sensitivity cardiac troponin depend on thresholds and time intervals
validated for a specific assay, yet these pathways are sometimes applied unchanged after a laboratory switches troponin
platforms. We review the assay-specific evidence behind the commonest 0/1-hour and 0/2-hour rule-out algorithms, the magnitude
of between-assay differences at the diagnostic decision limits these pathways use, and document a series of pathway-assay
mismatches identified across a hospital network following unannounced platform changes. Between-assay differences at the 99th
percentile upper reference limit ranged from 34% to 61% across four commonly used high-sensitivity troponin I and T assays, and
one network site continued applying a rule-out threshold validated for its previous assay for five months after switching platforms,
during which the pathway's negative predictive value could not be assured. We set out what a laboratory changing troponin assay
needs to revalidate, and propose a change-control checklist that ties any platform switch to a mandatory pathway threshold
review before the new assay goes live clinically.

pdf

References

Collet JP, Thiele H, Barbato E, et al. "2020 ESC guidelines for the management of acute coronary syndromes in patients presenting without persistent ST-segment elevation". Eur Heart J 42(2021): 1289-1367.

Apple FS, Sandoval Y, Jaffe AS, et al. "Cardiac troponin assays: guide to understanding analytical characteristics and their impact on clinical care". Clin Chem 63(2017): 73-81.

Wu AHB, Christenson RH, Greene DN, et al. "Clinical laboratory practice recommendations for the use of cardiac troponin in acute coronary syndrome". Clin Chem 64(2018): 645-655.

Aakre KM, Saenger AK, Body R, et al. "Analytical considerations in deriving 99th percentile upper reference limits for high-sensitivity cardiac troponin assays: educational recommendations". Clin Chem 68(2022): 1022-1030.

Clerico A, Zaninotto M, Passino C, et al. "Harmonization protocols for high-sensitivity cardiac troponin assays: how to put analytical and clinical requirements into practice". Clin Chem Lab Med 55(2017): 1636-1648.

Sandoval Y, Apple FS. "The global need to define normality: the 99th percentile value of cardiac troponin". Clin Chem 60(2014): 455-462.

Ordonez-Llanos J, Collinson PO, Christenson RH. "Amino-terminal pro-B-type natriuretic peptide: analytic considerations". Am J Cardiol 101(2008): 9-15.

Chapman AR, Adamson PD, Mills NL. "Assessment and classification of patients with myocardial injury and infarction in clinical practice". Heart 103(2017): 10-18.

Twerenbold R, Boeddinghaus J, Nestelberger T, et al. "Clinical use of high-sensitivity cardiac troponin in patients with suspected myocardial infarction". J Am Coll Cardiol 70(2017): 996-1012.

Aldous SJ, Richards AM, Cullen L, et al. "A 2-hour thrombolysis in myocardial infarction score outperforms other risk stratification tools in patients presenting with possible acute coronary syndromes". Acad Emerg Med 19(2012): 1128-1134.