Preanalytical Haemolysis in Emergency Department Samples: Rejection Rates and the Repeat Draw Nobody Counts
pdf

Keywords

Haemolysis
Preanalytical error
Specimen rejection
Emergency department
Laboratory quality

Abstract

Haemolysed samples are the commonest reason for specimen rejection in the emergency department, yet the repeat venepuncture
this forces is rarely tracked as an outcome in its own right. We measured haemolysis-related rejection across a twelve-month
period at three emergency departments, linked each rejection to whether a second draw was obtained and how long it took, and
examined which collection routes concentrated the problem. Haemolysis accounted for 61.4% of all preanalytical rejections
(2,014/3,281). Samples drawn through an existing intravenous cannula were rejected for haemolysis nearly three times as often as
samples drawn by direct venepuncture (9.8% vs 3.4% of draws, p<0.001), and the median time to a usable repeat sample was 47
minutes, rising to 74 minutes for patients triaged as time-critical. A targeted change in cannula-draw technique at one site cut its
haemolysis rejection rate by more than half within two quarters. Preanalytical haemolysis is not simply a laboratory quality
metric; it is a measurable source of diagnostic delay that responds to specific, identifiable changes in how samples are drawn.

pdf

References

Lippi G, Blanckaert N, Bonini P, et al. "Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories". Clin Chem Lab Med 46(2008): 764-772.

Heyer NJ, Derzon JH, Winges L, et al. "Effectiveness of practices to reduce blood sample haemolysis in EDs: a laboratory medicine best practices systematic review and meta-analysis". Clin Biochem 45(2012): 1012-1032.

Simundic AM, Bolenius K, Cadamuro J, et al. "Joint EFLM-COLABIOCLI recommendation for venous blood sampling". Clin Chem Lab Med 56(2018): 2015-2038.

Phelan MP, Reineks EZ, Schold JD, et al. "Preanalytic factors associated with hemolysis in emergency department blood samples". Arch Pathol Lab Med 142(2018): 229-235.

Cornes MP, Atherton J, Pourmahram G, et al. "Monitoring and reporting haemolysis in blood collection: best practice recommendations". Ann Clin Biochem 54(2017): 158-159.

Ong ME, Chan YH, Lim CS. "Reducing blood sample haemolysis at a tertiary hospital emergency department". Am J Med 121(2008): 1054-1059.

Kennedy C, Angermuller S, King R, et al. "A comparison of hemolysis rates using intravenous catheters versus venipuncture tubes for obtaining blood samples". J Emerg Nurs 22(1996): 566-569.

Halm MA. "Preventing hemolysis in blood samples drawn from vascular access devices". Am J Crit Care 21(2012): 210-213.

Lippi G, Cervellin G, Favaloro EJ, et al. "Preanalytical variability in laboratory testing: influence of the blood drawing technique". eJIFCC 30(2019): 90-104.

Green SF. "The cost of poor blood specimen quality and errors in preanalytical processes". Clin Biochem 46(2013): 1175-1179.