Turnaround time is one of the most widely reported laboratory quality indicators, yet the interval conventionally measured --
receipt of sample in the laboratory to result verification -- is not the interval a requesting clinician actually experiences, which
runs from the decision to order the test to the result appearing where the clinician will see it. We measured both intervals in
parallel across 14,206 emergency department chemistry orders at three hospitals over six months and quantified the gap between
them. The laboratory-only interval met the target of under sixty minutes for 91.4% of orders, while the clinician-experienced
interval, which additionally captured specimen collection delay and results system display latency, met the same sixty-minute
target for only 68.2% of the same orders. Specimen collection delay -- the time between order placement and sample arrival at the
bedside phlebotomy round -- accounted for the largest share of the gap. We argue that a laboratory reporting only the
conventional interval risks materially overstating service performance as experienced by the clinicians it serves.
References
Hawkins RC. "Laboratory turnaround time". Clin Biochem Rev 28(2007): 179-194.
Howanitz PJ. "Errors in laboratory medicine: practical lessons to improve patient safety". Arch Pathol Lab Med 129(2005): 1252-1261.
Steindel SJ, Howanitz PJ. "Physician satisfaction and emergency department laboratory test turnaround time". Arch Pathol Lab Med 125(2001): 863-871.
Goswami B, Singh B, Chawla R, et al. "Turn around time (TAT) as a benchmark of laboratory performance". Indian J Clin Biochem 25(2010): 376-379.
Lippi G, Da Rin G. "Notes for a proper laboratory turnaround time assessment in emergency medicine". Emerg Care J 6(2010): 1.
Fernandes CM, Worster A, Eva K, et al. "Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality". J Emerg Nurs 32(2006): 139-143.
Breil B, Fritz F, Thiemann V, et al. "Mapping turnaround times (TAT) to a generic timeline: a systematic review of TAT definitions in clinical domains". BMC Med Inform Decis Mak 11(2011): 34.
Hilborne LH, Oye RK, McArdle JE, et al. "Evaluation of stat and routine turnaround times as a component of laboratory quality". Am J Clin Pathol 91(1989): 331-335.
Novis DA, Walsh M, Wilkinson D, et al. "Continuous monitoring of stat and routine outlier turnaround times: two College of American Pathologists Q-Tracks monitors in 291 hospitals". Arch Pathol Lab Med 128(2004): 621-626.
Sanders DS. "Turnaround time - what does it mean, why does it seem so hard to achieve, and does it really matter anyway?". Am J Clin Pathol 133(2010): 173-174.

This work is licensed under a Creative Commons Attribution 4.0 International License.