Abstract
Small laboratories -- defined here as those processing fewer than 200 samples per analyte per month -- are reported anecdotally
to fail external quality assessment (EQA) surveys more often than larger laboratories, but the specific root causes behind these
failures are not well characterised separately from the failure rate itself. We reviewed root cause investigation records for every
EQA failure reported by 34 small laboratories across five consecutive survey cycles, categorising each investigated failure by its
identified root cause. Calibration or reagent lot-related causes accounted for the largest category (38.4%), followed by
insufficient sample volume for low-throughput analytes (24.1%) and infrequent operator exposure to a seldom-run assay (19.7%).
We argue that these three dominant categories share a common structural driver -- insufficient testing volume to detect drift or
maintain competence between EQA cycles -- and propose targeted mitigations for small laboratories that differ from the generic
EQA remediation guidance written primarily for higher-volume settings.
References
Miller WG, Jones GR, Horowitz GL, et al. "Proficiency testing/external quality assessment: current challenges and future directions". Clin Chem 57(2011): 1670-1680.
Sciacovelli L, Panteghini M, Lippi G, et al. "Defining a roadmap for harmonizing quality indicators in laboratory medicine: a consensus statement on behalf of the IFCC Working Group". Clin Chem Lab Med 55(2017): 1478-1488.
Libeer JC, Baadenhuijsen H, Fraser CG, et al. "Characterization and classification of external quality assessment schemes according to their purpose". Clin Chem Lab Med 34(1996): 665-678.
Jones GR, Albarede S, Kesseler D, et al. "Analytical performance specifications for external quality assessment - definitions and7 descriptions". Clin Chem Lab Med 55(2017): 949-955.
Sciacovelli L, Aita A, Padoan A, et al. "Performance criteria and quality indicators for the post-analytical phase". Clin Chem Lab Med 54(2016): 1169-1176.
Kristensen GB, Meijer P. "Interpretation of EQA results and EQA-based trouble shooting". Biochem Med 27(2017): 49-62.
Plebani M. "Quality indicators to detect pre-analytical errors in laboratory testing". Clin Biochem Rev 33(2012): 85-88.
Barth JH. "Selecting clinical quality indicators for laboratory medicine". Ann Clin Biochem 49(2012): 257-261.
Shahangian S, Snyder SR. "Laboratory medicine quality indicators: a review of the literature". Am J Clin Pathol 131(2009): 418-431.
Ricos C, Alvarez V, Cava F, et al. "Current databases on biological variation: pros, cons and progress". Scand J Clin Lab Invest 59(1999): 491-500.

This work is licensed under a Creative Commons Attribution 4.0 International License.